Specifications
582 Appendix D
Table of corresponding 4291B vs. E4991A GPIB commands
Table of Corresponding GPIB Commands
Table of Corresponding GPIB Commands
ABORt:
CALCulate subsystem:
4291B feature 4291B GPIB command E4991A GPIB command
Reset the trigger system and place all trigger sequences
in the idle state.
ABOR ABOR on page 286
4291B feature 4291B GPIB command E4991A GPIB command
Sets the partial search of the marker search function ON
or OFF.
CALC:EVAL:BAND:FULL CALC{1-5}:MARK:FUNC:DOM on
page 312
Sets the partial search range to the range between the
marker and the Δmarker.
CALC:EVAL:BAND:SPAN DMAR CALC{1-5}:MARK:FUNC:DOM:SP
AN on page 319
Sets the left (lower) border of the partial search range at
the current position of the marker.
CALC:EVAL:BAND:STAR MARK CALC{1-5}:MARK:FUNC:DOM:ST
AR on page 320
Sets the right (higher) border of the partial search range
at the current position of the marker.
CALC:EVAL:BAND:STOP MARK CALC{1-5}:MARK:FUNC:DOM:ST
OP on page 320
Sets the coupled or uncoupled marker mode. CALC:EVAL:COUP CALC:MARK:COUP on page 309
Selects the destination channel of the
marker->functions. When a marker->function is
performed, the sweep parameter or amplitude value of
the destination channel is changed.
CALC:EVAL:EFF:ON None (No concept of channel)
Calculates and display the equivalent circuit parameters. CALC:EVAL:EPAR CALC{1-5}:EPAR on page 295
Selects the equivalent circuit. CALC:EVAL:EPAR:CIRC CALC{1-5}:EPAR:CIRC on page 296
Sets the continuous or discontinuous marker mode. CALC:EVAL:INT CALC{1-5}:MARK:DISC on page 309
Queries the marker statistics. CALC:EVAL:MST:DATA? CALC{1-5}:MST:DATA? on page 344
Calculates the marker statistics in the portion of the
displayed trace that is in the search range.
CALC:EVAL:MST CALC{1-5}:MST on page 343
Displays the marker and selects the trace on which the
marker functions are used.
CALC:EVAL:ON1 "TR{1-21}" None (Among each of the GPIB
commands, specifies the trace number
that enables the marker function.)
Displays the voltage or current level applied to the DUT
at the marker point.
CALC:EVAL:ON2 CALC:BMON on page 291
(However, only the DC bias level can be
monitored.)
Sets the peak Δ value that is used to define the peak. CALC:EVAL:PEAK:EXC CALC{1-5}:MARK:APE:EXC:Y on
page 308
Sets the peak Δ value to the difference of amplitude
values between the preset marker position and both side
display points of the marker.
CALC{1-5}:MARK:APE:SET on
page 306
Sets the peak Δ value that is used to define the peak. CALC:EVAL:PEAK:EXC:X CALC{1-5}:MARK:APE:EXC:X on
page 307
Sets the peak Δ value to the difference of stimulus
values between the preset marker position and both side
display points of the marker.
CALC{1-5}:MARK:APE:SET on
page 306
Sets the peak polarity for the marker search functions. CALC:EVAL:PEAK:POL Chooses positive peak/negative peak in
the following command.
CALC{1-5}:MARK:FUNC on
page 311
Sets the threshold values. CALC:EVAL:PEAK:THR None (Cannot set the threshold of the
peak search function.)