Specifications

Chapter 17 523
GPIB Command Reference
SYST:KLOC:MOUS
17. GPIB Command
Reference
SYST:KLOC:MOUS
Format SYSTem:KLOCk:MOUSe {ON|OFF|1|0}
SYSTem:KLOCk:MOUSe?
Description Specifies whether to lock the mouse.
Parameters
Query response {1|0}<newline><^END>
Related commands SYST:KLOC command on page 522
SYST:KLOC:KBD command on page 522
Front Panel
Equivalents
Unavailable.
SYST:POFF
Format SYSTem:POFF
Description Turns E4991A's main power off. (No query)
Front Panel
Equivalents
Standby switch.
SYST:PRES
Format SYSTem:PRESet
Description Resets the E4991A to its initial settings. Resetting the E4991A by using the *RST
command provides basically the same function except for the points below.
Continuous initialization of the trigger system is set to ON.
Settings of data transfer format are not changed.
Related commands *RST command on page 282
Front Panel
Equivalents
[Preset] key
Description
ON or 1 Mouse locked
OFF or 0 (Default) Mouse unlocked