Specifications
Chapter 17 515
GPIB Command Reference
SWE:TYPE
17. GPIB Command
Reference
SWE:TYPE
Format [SENSe:]SWEep:TYPE {LINear|LOGarithmic|SEGMent|POWer|DCV|DCI}
[SENSe:]SWEep:TYPE?
Description Selects the sweep parameter (sweep type).
Parameters
Query response {LIN|LOG|SEGM|POW|DCV|DCI}<newline><^END>
Front Panel
Equivalents
Stimulus - Sweep Setup... - Sweep Parameter|Sweep Type
Description
LINear (Default) Specifies the frequency sweep (linear sweep).
LOGaritmic Specifies the frequency sweep (log sweep).
SEGMent
*1
Specifies the frequency sweep (segment sweep).
POWer Specifies the oscillator level sweep (linear sweep).
DCV
*2
Specifies the dc bias voltage sweep (linear sweep).
DCI
*2
Specifies the dc bias current sweep (linear sweep).
*1.In order to set the segment sweep, it is necessary to create the segment sweep table in
advance.
*2.This can be used when Option 001 (dc bias function) is installed.