Specifications

508 Chapter 17
GPIB Command Reference
SWE:DIR
SWE:DIR
Format [SENSe:]SWEep:DIRection {UP|DOWN}
[SENSe:]SWEep:DIRection?
Description Selects the direction of the sweep.
Parameters
Query response {UP|DOWN}<newline><^END>
Front Panel
Equivalents
Stimulus - Sweep Setup... - Sweep Direction: [Up/Down]
SWE:DWEL1
Format [SENSe:]SWEep:DWELl1 <numeric>
[SENSe:]SWEep:DWELl1?
Description Sets the delay time for each sweep.In addition, when automatic sweep time is set to auto
with the
SWE:TIME:AUTO command, the delay time is reset to 0 second.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SWE:TIME:AUTO command on page 514
Front Panel
Equivalents
Stimulus - Sweep Setup... - Sweep Time: [Auto/Manual] - Sweep Delay
Description
UP (Default) Specifies the direction toward the increasing stimulus value.
DOWN Specifies the direction toward the decreasing stimulus value.
<numeric>
Description Delay time
Data type Floating point
Range 0 to 20
Default 0
Resolution 1E-4
Unit s (seconds)