Specifications

488 Chapter 17
GPIB Command Reference
SOUR:VOLT:STOP
SOUR:VOLT:STOP
Format SOURce:VOLTage:STOP <numeric>
SOURce:VOLTage:STOP?
Description Sets the stop value in the oscillator voltage level sweep range.
In addition, use the SOUR:VOLT:STAR command to set the start value in the sweep
range.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SOUR:VOLT:MODE command on page 479
SOUR:VOLT:STAR command on page 487
Front Panel
Equivalents
Stimulus - Start/Stop... - Stop: V
<numeric>
Description Stop value in the sweep range
Data type Floating point
Range 5E-3 to 502E-3
Default 400E-3
Resolution 1E-3
Unit V (volt)