Specifications

Chapter 17 487
GPIB Command Reference
SOUR:VOLT:STAR
17. GPIB Command
Reference
SOUR:VOLT:STAR
Format SOURce:VOLTage:STARt <numeric>
SOURce:VOLTage:STARt?
Description Sets the start value in the oscillator voltage level sweep range.
In addition, use the SOUR:VOLT:STOP command to set the stop value in the sweep
range.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SOUR:VOLT:MODE command on page 479
SOUR:VOLT:STOP command on page 488
Front Panel
Equivalents
Stimulus - Start/Stop... - Start: V
<numeric>
Description Start value in the sweep range
Data type Floating point
Range 5E-3 to 502E-3
Default 200E-3
Resolution 1E-3
Unit V (volt)