Format SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]:OFFSet:STARt <numeric>
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]:OFFSet:STARt?
Description Sets the start value in the dc bias voltage sweep range. This command can be used when
Option 001 (dc bias function) is installed.
In addition, use the SOUR:VOLT:OFFS:STOP command to set the stop value of the
sweep range.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SOUR:VOLT:OFFS:STOP command on page 485
Front Panel
Equivalents
Stimulus - Start/Stop... - Start: V
<numeric>
Description Start value in the sweep range
Data type Floating point
Range -40 to 40
Default 0
Resolution 1E-3
Unit V (volt)