Specifications

456 Chapter 17
GPIB Command Reference
SENS:CORR2:FIXT:EDEL:USER:DIST
SENS:CORR2:FIXT:EDEL:USER:DIST
Format SENSe:CORRection2:FIXTure:EDELay:USER:DISTance <numeric>
SENSe:CORRection2:FIXTure:EDELay:USER:DISTance?
Description Sets the electrical length that compensates for the delay caused by the connection of a
custom test fixture. If you want to configure the instrument for your custom test fixture
connected to the DUT PORT, use this command to set an electrical length that can
compensate for the resulting delay.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SENS:CORR2:FIXT command on page 454
Front Panel
Equivalents
Stimulus - Cal/Comp... - Fixture Length: (m)
<numeric>
Description Electrical length
Data type Floating point
Range -1E3 to 1E3
Default 0
Unit m (meter)