Specifications
Chapter 17 451
GPIB Command Reference
SENS:CORR2:COLL:OPEN
17. GPIB Command
Reference
SENS:CORR2:COLL:OPEN
Format SENSe:CORRection2:COLLect:OPEN[:STATe] {ON|OFF|1|0}
SENSe:CORRection2:COLLect:OPEN[:STATe]?
Description Turns on/off the Open compensation function. At this time, the fixture compensation
coefficient is re-calculated. If this command is executed before measuring the Open
compensation data with the
SENS:CORR2:COLL command, an error will occur and the
command will be ignored.
Parameters
Query response {1|0}<newline><^END>
Related commands SENS:CORR2:COLL command on page 449
Front Panel
Equivalents
Stimulus - Cal/Comp... - Comp Menu - Comp Open [On/Off]
SENS:CORR2:COLL:SAVE
Format SENSe:CORRection2:COLLect:SAVE
Description Calculates the fixture compensation coefficient and sets the fixture compensation function
to on. If this command is executed before measuring the fixture compensation data with the
SENS:CORR2:COLL command, an error will occur and the command will be ignored.
(No query)
Related commands SENS:CORR2:COLL command on page 449
Front Panel
Equivalents
Stimulus - Cal/Comp... - Comp Menu - Done
Description
ON or 1 Turns on the Open compensation function.
OFF or 0 (Default) Turns off the Open compensation function.