Specifications

446 Chapter 17
GPIB Command Reference
SENS:CORR2:CKIT:STAN2:LIST:R
SENS:CORR2:CKIT:STAN2:LIST:R
Format SENSe:CORRection2:CKIT:STANdard2:LIST:R <numeric 1>,<numeric 2>,..,<numeric
N>
SENSe:CORRection2:CKIT:STANdard2:LIST:R?
Description Defines the resistance value (R) of the Short standard of the fixture compensation kit by
using the list setting function.
When using the list setting function, the resistance value (R) of the Short standard, defined
for the calibration data throughout the entire frequency range, is set for the exact number of
measurement points (N).
NOTE The standard value defined by the list setting function can be used when measuring the
fixture compensation data in the user-defined frequency point/user-defined power point.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric 1}, {numeric 2},.., {numeric N}<newline><^END>
Where N is the number of measurement points.
Related commands SENS:CORR2:CKIT:LIST command on page 440
SENS:CORR2:CKIT:STAN2:LIST:X command on page 447
FORM:DATA command on page 374
Front Panel
Equivalents
Unavailable.
<numeric n>
Description Resistance value (R) of the Short standard
Data type Floating point
Range -1E6 to 1E6
Default 0
Unit Ω (ohm)