Specifications

444 Chapter 17
GPIB Command Reference
SENS:CORR2:CKIT:STAN1:LIST:G
SENS:CORR2:CKIT:STAN1:LIST:G
Format SENSe:CORRection2:CKIT:STANdard1:LIST:G <numeric 1>,<numeric 2>,..,<numeric
N>
SENSe:CORRection2:CKIT:STANdard1:LIST:G?
Description Defines the conductance value (G) of the Open standard of the fixture compensation kit by
using the list setting function..
When using the list setting function, the conductance value (G) of the Open standard,
defined for the calibration data throughout the entire frequency range, is set for the exact
number of measurement points (N).
NOTE The standard value defined by the list setting function can be used when measuring the
fixture compensation data in the user-defined frequency point/user-defined power point.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric 1}, {numeric 2},.., {numeric N}<newline><^END>
Where N is the number of measurement points.
Related commands SENS:CORR2:CKIT:LIST command on page 440
SENS:CORR2:CKIT:STAN1:LIST:B command on page 443
FORM:DATA command on page 374
Front Panel
Equivalents
Unavailable.
<numeric n>
Description Conductance value (G) of the Open standard
Data type Floating point
Range -1E6 to 1E6
Default 0
Unit S (Siemens)