Specifications

436 Chapter 17
GPIB Command Reference
SENS:CORR1:CKIT:STAN7:THIC
SENS:CORR1:CKIT:STAN7:THIC
Format SENSe:CORRection1:CKIT:STANdard7:THICkness <numeric>
SENSe:CORRection1:CKIT:STANdard7:THICkness?
Description Defines the thickness of the Load standard of the calibration kit in dielectric measurement.
This command can be used when Option 002 (material measurement software) is installed.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SENS:CORR1:CKIT:STAN7:PLF command on page 434
SENS:CORR1:CKIT:STAN7:PRE command on page 435
Front Panel
Equivalents
Stimulus - Cal/Comp... - Cal Kit Menu - Thickness
<numeric>
Description Thickness of the Load standard
Data type Floating point
Range -1E6 to 1E6
Default 800E-6
Unit m (meter)