Specifications

430 Chapter 17
GPIB Command Reference
SENS:CORR1:CKIT:STAN3:L
SENS:CORR1:CKIT:STAN3:L
Format SENSe:CORRection1:CKIT:STANdard3:L <numeric>
SENSe:CORRection1:CKIT:STANdard3:L?
Description Defines the inductance value (L) of the Load standard of a user-defined calibration kit.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SENS:CORR1:CKIT:LIST command on page 421
SENS:CORR1:CKIT:STAN3:R command on page 433
Front Panel
Equivalents
Stimulus - Cal/Comp... - Cal Kit Menu - Load L: (H)
<numeric>
Description Inductance value (L) of the Load standard
Data type Floating point
Range -1E6 to 1E6
Default 0
Unit H