Specifications

Chapter 3 43
Setting Measurement Conditions
Set Measurement Parameter
3. Setting Measurement
Conditions
Parameters for Complex Trace
Measurement parameters
Table 3-2 shows measurement parameters available for a complex trace. Use the following
GPIB command to specify a trace number (4 or 5) and select measurement parameters.
Note that up to two measurement parameters can be displayed at any one time.
CALC{1-5}:FORM on page 298
Display Format
Table 3-3 shows display formats available for a complex trace. Use the following GPIB
command to select a display format from Complex plane format, Polar format, Smith chart,
or Admittance chart.
DISP:TRAC{1-5}:GRAT:FORM on page 357
Table 3-2 Measurement Parameters for Complex Trace
Measurement
parameter
GPIB
parameter
Description
Z Z Impedance
Y Y Admittance
Γ RC Reflection coefficient
μr P
Complex permeability
*1
εr DC
Complex permittivity
*2
*1. Parameters selectable when the Magnetic Material measurement mode is selected.
*2. Parameters selectable when the Dielectric Material measurement mode is selected.
Table 3-3 Display Format for Complex Trace
Measurement
parameters
Display format
Z,Y,μr,εr Complex plane or Polar format
Γ Complex plane format, Polar format, Smith chart, or Admittance
chart