Specifications
Chapter 17 415
GPIB Command Reference
SEGM{1-16}:VOLT:LIM
17. GPIB Command
Reference
SEGM{1-16}:VOLT:LIM
Format [SENSe:]SEGMent{1-16}:VOLTage:LIMit <numeric>
[SENSe:]SEGMent{1-16}:VOLTage:LIMit?
Description When creating the segment sweep table, sets the dc bias voltage limit maximum value.
This command is available when Option 001 (dc bias function) is installed.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SEGM:CURR:OFFS:STAT command on page 399
SEGM{1-16}:CURR:OFFS command on page 401
Front Panel
Equivalents
Stimulus - Sweep Setup... - Segment Table Menu - More - Bias Limit [V]
Sub-block Description
SEGMent {1-16} Specifies the segment number (1 to 16).
<numeric>
Description dc bias voltage limit maximum value
Data type Floating point
Range 1 to 40
Default 1
Resolution 1E-3
Unit V (volt)