Specifications

414 Chapter 17
GPIB Command Reference
SEGM{1-16}:VOLT
SEGM{1-16}:VOLT
Format [SENSe:]SEGMent{1-16}:VOLTage[:LEVel] <numeric>
[SENSe:]SEGMent{1-16}:VOLTage[:LEVel]?
Description When creating the segment sweep table, sets the oscillator voltage level.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SEGM:VOLT:STAT command on page 418
Front Panel
Equivalents
Stimulus - Sweep Setup... - Segment Table Menu - More - Osc Level: V
Sub-block Description
SEGMent {1-16} Specifies the segment number (1 to 16).
<numeric>
Description Oscillator voltage level
Data type Floating point
Range 5E-3 to 502E-3
Default 100E-3
resolution 1E-3
Unit V (volt)