Specifications

412 Chapter 17
GPIB Command Reference
SEGM:POW:STAT
SEGM:POW:STAT
Format [SENSe:]SEGMent:POWer:STATe {ON|OFF|1|0}
[SENSe:]SEGMent:POWer:STATe?
Description When creating the segment sweep table, selects the power mode to set the oscillator level.
In addition, use the SEGM{1-16}:POW command to set the oscillator power level.
Parameters
Query response {1|0}<newline><^END>
Related commands SEGM{1-16}:POW command on page 411
SEGM:CURR:STAT command on page 402
SEGM:VOLT:STAT command on page 418
Front Panel
Equivalents
Stimulus - Sweep Setup... - Segment Table Menu - More - Osc Unit
Description
ON or 1 Specifies the power (dBm) mode.
OFF or 0 (Default)
Specifies a mode other than power mode.
*1
*1.When specifying a mode other than the power mode, specify either current mode or
voltage mode with the SEGM:CURR:STAT command or the SEGM:VOLT:STAT
command, respectively.