Specifications
Chapter 17 407
GPIB Command Reference
SEGM{1-16}:FREQ:CENT
17. GPIB Command
Reference
SEGM{1-16}:FREQ:CENT
Format [SENSe:]SEGMent{1-16}:FREQuency:CENTer <numeric>
[SENSe:]SEGMent{1-16}:FREQuency:CENTer?
Description When creating the segment sweep table, sets the center value in the frequency sweep range.
In addition, use the SEGM{1-16}:FREQ:SPAN command when setting the span value in
the sweep range.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SEGM{1-16}:FREQ:SPAN command on page 408
Front Panel
Equivalents
Unavailable.
Sub-block Description
SEGMent {1-16} Specifies the segment number (1 to 16).
<numeric>
Description Center value in the sweep range
Data type Floating point
Range 1E6 to 3E9
Default 1.5005E9
Resolution 1E-3
Unit Hz