Specifications
Chapter 17 403
GPIB Command Reference
SEGM{1-16}:DATA
17. GPIB Command
Reference
SEGM{1-16}:DATA
Format [SENSe:]SEGMent{1-16}:DATA <numeric 1>,<numeric 2>,<numeric 3>,<numeric
4>,<numeric 5>,<numeric 6>,<numeric 7>
Description When creating/editing the segment sweep table, simultaneously sets the parameter values
of all segments. Before executing this GPIB command, you need to set the oscillator level
to be read as current, dBm, or voltage by using the
SEGM:CURR:STAT command,
SEGM:POW:STAT command, or SEGM:VOLT:STAT command, respectively. When
applying dc bias, you need to set either current or voltage by using the
SEGM:CURR:OFFS:STAT command or SEGM:VOLT:OFFS:STAT command,
respectively.
Parameters
Sub-block Description
SEGMent {1-16} Specifies the segment number (1 to 16).
<numeric 1> <numeric 2> <numeric 3> <numeric 4>
Description Start
frequency
Stop
frequency
Number of
points
Point
averaging
factor
Data type Floating point Integer
Range 1E6 to 3E9 1E6 to 3E9
2 to 801
*1
1 to 100
Default 1E6 3E9 2 1
Unit Hz None
*1.The number of points cannot be set beyond 801 points, which is the sum of the points in
all segments.
<numeric 5> <numeric 6> <numeric 7>
Description Oscillator level dc bias
current/voltage
level
dc bias
voltage/current
limit maximum
value
Data type Floating point
Range
voltage:
5E-3 to 502E-3
current:
0.1E-3 to 10E-3
power:
-40 to 1
voltage:
-40 to 40
current:
-50E-3 to -100E-6,
100E-6 to 50E-3
current:
2E-3 to 50E-3
voltage:
1 to 40