Specifications

402 Chapter 17
GPIB Command Reference
SEGM:CURR:STAT
SEGM:CURR:STAT
Format [SENSe:]SEGMent:CURRent:STATe {ON|OFF|1|0}
[SENSe:]SEGMent:CURRent:STATe?
Description When creating the segment sweep table, selects the current mode to set the oscillator level.
In addition, use the SEGM{1-16}:CURR command to set the oscillator current level.
Parameters
Query response {1|0}<newline><^END>
Related commands SEGM{1-16}:CURR command on page 398
SEGM:POW:STAT command on page 412
SEGM:VOLT:STAT command on page 418
Front Panel
Equivalents
Stimulus - Sweep Setup... - Segment Table Menu - More - Osc Unit
Description
ON or 1 Specifies the current mode.
OFF or 0 (Default)
Specifies the mode other than the current mode.
*1
*1.When specifying a mode other than the current mode, specify either the power mode or
the voltage mode with the SEGM:POW:STAT command or SEGM:VOLT:STAT
command.