Specifications

Chapter 17 399
GPIB Command Reference
SEGM:CURR:OFFS:STAT
17. GPIB Command
Reference
SEGM:CURR:OFFS:STAT
Format [SENSe:]SEGMent:CURRent:OFFSet:STATe {ON|OFF|1|0}
[SENSe:]SEGMent:CURRent:OFFSet:STATe?
Description When creating the segment sweep table, selects the voltage mode to set the dc bias level.
This command can be used when Option 001 (dc bias function) is installed.
In addition, use the SEGM{1-16}:CURR:OFFS command to select the current mode to
set the dc bias level.
Parameters
Query response {1|0}<newline><^END>
Related commands SEGM{1-16}:CURR:OFFS command on page 401
SEGM{1-16}:VOLT:LIM command on page 415
Front Panel
Equivalents
Stimulus - Sweep Setup... - Segment Table Menu - More - Bias Source
Description
ON or 1 (Default) Specifies the current mode.
OFF or 0 Specifies the voltage mode.