Specifications
398 Chapter 17
GPIB Command Reference
SEGM{1-16}:CURR
SEGM{1-16}:CURR
Format [SENSe:]SEGMent{1-16}:CURRent[:LEVel] <numeric>
[SENSe:]SEGMent{1-16}:CURRent[:LEVel]?
Description When creating the segment sweep table, sets the oscillator current level.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands SEGM:CURR:STAT command on page 402
Front Panel
Equivalents
Stimulus - Sweep Setup... - Segment Table Menu - More - Osc Level: A
Sub-block Description
SEGMent {1-16} Specifies the segment number (1 to 16).
<numeric>
Description Oscillator current level
Data type Floating point
Range 0.1E-3 to 10E-3
Default 2E-3
Resolution 0.01E-3
Unit A (ampere)