Specifications

Chapter 17 377
GPIB Command Reference
FREQ:SPAN
17. GPIB Command
Reference
FREQ:SPAN
Format [SENSe:]FREQuency:SPAN <numeric>
[SENSe:]FREQuency:SPAN?
Description Sets the span value in the frequency sweep range (linear/log sweep).
In addition, use the FREQ:CENT command to set the center value in the sweep range.
Parameters
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Related commands FREQ:CENT command on page 376
Front Panel
Equivalents
Stimulus - Start/Stop... - Span: Hz
FREQ:SPAN:FULL
Format [SENSe:]FREQuency:SPAN:FULL
Description Sets the sweep range in full span (1 M to 3 GHz) for the frequency sweep (linear/log
sweep). (No query)
Front Panel
Equivalents
Unavailable.
<numeric>
Description Span value in the sweep range
Data type Floating point
Range 1E6 to 2999E6
Default 2.999E9
Resolution 1E-3
Unit Hz