Specifications
Chapter 17 375
GPIB Command Reference
FREQ
17. GPIB Command
Reference
FREQ
Format [SENSe:]FREQuency[:CW|:FIXed] <numeric>
[SENSe:]FREQuency[:CW|:FIXed]?
Description Sets the CW frequency for the oscillator level sweep, dc bias voltage sweep, or dc bias
current sweep.
Parameters Either CW or FIXed can be written when describing the sub-block of the command,
although [:CW|:FIXes] in the sub-block part can be omitted.
If the specified parameter is outside the range, the minimum (if the lower limit of the range
is exceeded) or maximum value (if the upper limit is exceeded) will be assumed.
Query response {numeric}<newline><^END>
Front Panel
Equivalents
Stimulus - Source... - CW Freq: Hz
<numeric>
Description CW frequency
Data type Floating point
Range 1E6 to 3E9
Default 1E6
Resolution 1E-3
Unit Hz