Specifications

372 Chapter 17
GPIB Command Reference
DISP:TRAC{1-3}:Y:TOP
DISP:TRAC{1-3}:Y:TOP
Format DISPlay[:WINDow]:TRACe{1-3}:Y[:SCALe]:TOP <numeric>
DISPlay[:WINDow]:TRACe{1-3}:Y[:SCALe]:TOP?
Description Sets the maximum scale value in the Y-axis direction for the scalar trace.
In addition, use the DISP:TRAC{1-3}:Y:BOTT command to set the minimum scale
value.
Parameters
Query response {numeric}<newline><^END>
Related commands DISP:TRAC{1-3}:Y:BOTT command on page 365
Front Panel
Equivalents
Scale - Scale... - Top
Sub-block Description
TRACe{1-3} Specifies the trace number (1 to 3).
<numeric>
Description Maximum scale value in the Y-axis direction
Data type Floating point
Default Variable depending on the measurement parameter.
1E-3 Cs/Cp
1 B/D/tanδ (ε)/tanδ (μ)/G/|Γ|/Γx/Γy/|Y|
10 Ls/Lp
200 θγ/θz/θy
1E3 Q
1E6 |εr|/εr'/εr"/|μr|/μr'/μr"/R/Rs/Rp/|Z|/X
Unit Variable depending on the measurement parameter