Specifications

368 Chapter 17
GPIB Command Reference
DISP:TRAC{1-5}:Y:PDIV
Related commands DISP:TRAC{1-5}:Y:PDIV command on page 368
Front Panel
Equivalents
Scale - Scale... - Full-scale|Scale
DISP:TRAC{1-5}:Y:PDIV
Format DISPlay[:WINDow]:TRACe{1-5}:Y[:SCALe]:PDIVision <numeric>
DISPlay[:WINDow]:TRACe{1-5}:Y[:SCALe]:PDIVision?
Description Sets the Y-axis direction scale by using a one-tenths value of full-scale. However, the
setting value here does not always have a one-to-one correspondence to the scale value per
division.
In addition, use the DISP:TRAC{1-5}:Y:FULL command to set the full-scale value.
Parameters
Query response {numeric}<newline><^END>
Related commands DISP:TRAC{1-5}:Y:FULL command on page 367
Front Panel
Equivalents
Unavailable.
Sub-block Description
CALCulate{1-5} Specifies the trace number (1 to 5).
<numeric>
Description One-tenths value of full-scale
Data type Floating point
Range Variable depending on the measurement parameter.
In the Linear Y-axis format:
1E-4 Cs/Cp
0.1 B/D/tanδ (ε)/tanδ (μ)/G/|Y|
0.2 |Γ|/Γx/Γy
1 Ls/Lp
36 θγ/θz/θy
1E2 Q
1E5 |εr|/εr'/εr"/|μr|/μr'/μr"/R/Rs/Rp/|Z|/X
In the Complex format:
50 Z/Y/Γ/εr/μr
Default Variable depending on the measurement parameter.