Specifications

336 Chapter 17
GPIB Command Reference
CALC{1-5}:MARK:SET
CALC{1-5}:MARK:SET
Format CALCulate{1-5}:MARKer:SET
{CENTer|DELTASPAN|STARt|STOP|REFerence|OFFSet}
Description Performs the E4991A’s settings by using the stimulus/measurement value at the active
marker positions. (No query)
Parameters
Front Panel
Equivalents
Marker - To... - Center|DeltaToSpan|Start|Stop|Reference|Offset
Sub-block Description
CALCulate{1-5} Specifies the trace number (1 to 5).
Description
CENTer Sets the stimulus value at the active marker position to the center
value in the sweep range.
DELTASPAN
*1
Sets the area where the active marker and reference marker are
placed to the span value in the sweep range.
STARt Sets the stimulus value at the active marker position to the start
value in the sweep range.
STOP Sets the stimulus value at the active marker position to the stop
value in the sweep range.
REFerence
*2
Sets the measurement value at the active marker position to the
reference value on the Y-axis.
OFFSet
*3
Sets the measurement value at the active marker position to the
offset value to be subtracted from the trace data.
*1.Can be set when the Δ mode is set to on.
*2.Can be set when the display format is the linear Y-axis format or the complex plane for-
mat.
*3.Can be set for the scalar trace.