Specifications
Chapter 17 335
GPIB Command Reference
CALC{1-5}:MARK:REF:Y
17. GPIB Command
Reference
CALC{1-5}:MARK:REF:Y
Format CALCulate{1-5}:MARKer:REFerence:Y <numeric 1>,<numeric 2>
CALCulate{1-5}:MARKer:REFerence:Y?
Description Moves the reference marker (marker R) to the specified measurement value when the Δ
mode is set to fixed Δ. In addition, use the
CALC{1-5}:MARK:REF:X command to move
the reference marker position to the specified stimulus value.
Parameters • With scalar trace:
• With complex trace:
Query response {numeric 1}, {numeric 2}<newline><^END>
Related commands CALC{1-5}:MARK:REF:TYPE command on page 333
CALC{1-5}:MARK{1-8}:X command on page 338
Front Panel
Equivalents
Marker - Marker... - Delta Marker Menu - Delta Value|Delta Aux Value
Sub-block Description
CALCulate{1-3} Specifies the trace number (1 to 3).
<numeric 1> <numeric 2>
Description Measurement value in the real
number type
Always 0
Data type Floating point
Range -1E12 to 1E12
Default Variable depending on the sweep
parameter.
0
Unit None
Sub-block Description
CALCulate{4-5} Specifies the trace number (4 to 5).
<numeric 1> <numeric 2>
Description Real part of measurement value
(complex type)
Imaginary part of measurement
value (complex type)
Data type Floating point
Range -1E12 to 1E12
Default Variable depending on the sweep parameter.
Unit