Specifications
Chapter 17 333
GPIB Command Reference
CALC{1-5}:MARK:REF:TYPE
17. GPIB Command
Reference
CALC{1-5}:MARK:REF:TYPE
Format CALCulate{1-5}:MARKer:REFerence:TYPE {OFF|DELTA|FIXDELTA}
CALCulate{1-5}:MARKer:REFerence:TYPE?
Description Selects the type of reference marker (marker R).
Parameters
Query response {OFF|DELTA|FIXDELTA}<newline><^END>
Front Panel
Equivalents
Marker - Marker... - Delta Marker Menu - Delta Mode
Sub-block Description
CALCulate{1-5} Specifies the trace number (1 to 5).
Description
OFF (Default) Sets the Δ mode to off.
DELTA Sets the Δ mode to on. At this time, the reference marker is fixed
at the position of the current stimulus value. The reference
marker moves in accordance with the measurement value
variation of the specified stimulus value during the measurement.
FIXDELTA Sets the Δ mode to on. At this time, the reference marker is fixed
at the position of the current stimulus value and measurement
value. The reference marker is fixed during the measurement
regardless of the measurement value variation.