Specifications
Chapter 17 295
GPIB Command Reference
CALC{1-5}:EPAR
17. GPIB Command
Reference
CALC{1-5}:EPAR
Format CALCulate{1-5}:EPARameters
Description Executes the equivalent circuit analysis in the selected equivalent circuit model. In
addition, executes the equivalent circuit analysis in the marker partial search range. When
the marker partial search range is not set to on with the
CALC{1-5}:MARK:FUNC:DOM
command, executes the equivalent circuit analysis in the entire sweep range. (No query)
In addition, use the CALC{1-5}:EPAR:CIRC command to select the equivalent circuit
mode.
Also, the equivalent circuit parameters C0, C1, L1, and R1 that were acquired in the
equivalent circuit analysis can be read when executing the
CALC{1-5}:DATA:EPAR
command in the query.
Parameters
Related commands CALC{1-5}:EPAR:CIRC command on page 296
CALC{1-5}:DATA:EPAR command on page 294
Front Panel
Equivalents
Utility - Equivalent Circuit... - Calculate Parameters
Sub-block Description
CALCulate{1-5} Specifies the trace number (1 to 5).