Measurement Start and Detection of Measurement End
Measurement Trigger (Measurement Start)
Measurement Trigger (Measurement Start)
Trigger System
The E4991A trigger system has three states: “Idle,” “Waiting for trigger,” and
“Measuring.” as shown in
Figure 5-1.
Figure 5-1 Trigger System
The following section explains the state transitions of the trigger system.
Idle State
Immediately after the E4991A is powered on, the trigger system is in the Idle state.
However, the trigger system soon transits to the Waiting for trigger state because the
trigger system’s continuous initiation feature is enabled and the trigger source is set to
internal trigger. After that, the E4991A transits between the Measuring and Waiting for
trigger states. Also, when one of the following GPIB commands is executed, the trigger
system transits to the Idle state.
• *RST on page 282
• ABOR on page 286
When the dc bias function has been switched from disabled to enabled, the trigger system
transits to the Idle state.
If the trigger system has been initiated using the following GPIB command, the state
changes from Idle to Waiting for trigger.
• INIT on page 382
• INIT:CONT on page 382