Technical data
62 Agilent Connectivity Guide
3 Connecting Instruments to GPIB
GPIB Hardware Checks
This section gives guidelines to make hardware troubleshooting checks
for the GPIB card (such as an Agilent 82351 PCIe™-GPIB), including:
• Check Cables/Connections/Power.
• Check Device Manager
• If Sound Card Does Not Work
• If Data Transfers to Devices Fail
• If Connection Expert Finds an Agilent 82350 Card with Serial
Number ffffffff
Check Cables/Connections/Power. Start your troubleshooting
sequence by performing the following hardware checks. If the hardware
checks do not solve the problem, see GPIB Software Checks.
1
Check GPIB Cable Connections. Check all GPIB cables for good
connection to the GPIB connector on the interface card installed in
your PC and the GPIB cable connections between all connected
GPIB instruments. An improperly attached GPIB connector can
cause the bus to malfunction.
2
Check GPIB Cables for Damage. Check all GPIB cables for
cuts/damage and check for bent/misaligned/crushed connector pins.
Replace cables as required.
3
Disconnect/Reconnect GPIB Cables. If Steps 1 and 2 do not solve the
problem, try disconnecting and reconnecting (or replacing) GPIB
cables.
4
Check PC/Instrument Power-on. Verify that the PC and all connected
GPIB instruments are functional and are powered ON. Verify that
host computer is not in a Suspended power management state.
5
Reboot Your PC. If doing Steps 1, 2, 3 or 4 does not solve the
problem, reboot the PC. If this does not solve the problem, go to
Check Device Manager.
NOTE
There are no user-serviceable parts for the GPIB Interface cards. If you
suspect a hardware failure for interface card, contact Agilent for
instructions to replace or repair the unit. See Contacting Agilent for
telephone numbers/web site address.