User`s manual

Relay Matrix Switch Error Messages 77Appendix C
Appendix C
Relay Matrix Switch Error Messages
This table lists the error messages associated with the Relay Matrix Switch
modules when programmed with SCPI. See the appropriate command
module user’s manual for complete information on error messages.
Number Title Potential Cause(s)
-211 Trigger ignored Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.
-213 Init Ignored Attempting to execute an INIT command when a scan is already in
progress.
-224 Illegal parameter value Attempting to execute a command with a parameter not applicable to the
command.
-350 Too many errors. The queue holds a maximum of 30 error numbers/messages for each
switchbox. The queue has overflowed.
1500 External trigger source
already allocated
Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another switchbox.
2000 Invalid card number Addressing a module (card) in a switchbox that is not part of the
switchbox.
2001 Invalid channel number Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g.,, channel 99 of a multiplexer module).
2006 Command not supported
on this card
Sending a command to a module (card) in a switchbox that is
unsupported by the module.
2008 Scan list not initialized Executing a scan without the INIT command.
2009 Too many channels in
channel list
Attempting to address more channels than available in the switchbox.
2012 Invalid Channel Range Invalid channel(s) specified in SCAN <
channel_list> command.
Attempting to begin scanning when no valid channel list is defined.
2600 Function not supported on
this card
Sending a command to a module (card) in a switchbox that is not
supported by the module or switchbox.
2601 Channel list required Sending a command requiring a channel list without the channel list.