User`s manual
Getting Started 21Chapter 1
E1469A Terminal Module
Connectors
Figure 1-9 shows the E1469A terminal module connectors and associated
row/column designators. Shielding jumpers JM1 - JM12 are shown. See
"Creating Larger Matrices" for information on using the expansion
connectors J1 - J5 and for shield wiring details.
NOTE Jumpers JM1 - JM12 on the E1469A terminal module connect row/column
shields to earth ground through the VXIbus backplane. You may want to
remove one or more of these jumpers to reduce common mode noise.
Available Cables To assist you in wiring Relay Matrix Switch terminal modules into your test
system, this table shows a list of cables that are available from Agilent.
Figure 1-9. E1469A 4 x 16 Matrix Switch Terminal Block
Connector
Row
Column Expansion
Connectors
*
Column Input
Connectors
Row Input
Connector
Row Expansion
Connector
Shield
Connector
TB5
***
In parallel with the screw terminals.
*
Connectors
Column
Description Finished
Length
End "A" End "B" Part Number
Module expansion connector
with quick disconnect
(twisted pair)
~30 cm 4 x 2 connector for
expansion connectors on
terminal modules
4 x 2 connector for
expansion connectors on
terminal modules
E1468-80002
50 Coax 2.0 m 2-pin TLA* BNC (molded over) E1065-61620
Dual banana instrument 2.0 m 3-pin TLA* Dual banana E1066-61620
SMB instrument 2.0 m 2-pin TLA* SMB (molded over) E1068-61620
*TLA is a family of connector/cable assemblies with good transmission line design that are made by an Agilent supplier.
The 2-pin and 3-pin TLA connectors are designed to fit on one channel of the terminal module expansion connectors.