User`s manual

20 Getting Started Chapter 1
Configuring the Terminal Modules
This section gives guidelines to configure the E1468A and E1469A terminal
modules, including:
Wiring the Terminal Module
Creating Larger Matrixes
Attaching a Terminal Module to the Relay Switch Module
Wiring the Terminal
Modules
Guidelines to wire the E1468A and E1469A terminal modules follow.
E1468A Terminal Module
Connectors
Figure 1-8 shows the E1468A terminal module connectors and associated
row/column designators. Shielding jumpers JM1 - JM10 are shown. See
"Creating Larger Matrices" for information on using the expansion
connectors J1 - J4 and for shield wiring details.
NOTE Jumpers JM1 - JM10 on the E1468A terminal module connect row/column
shields to earth ground through the VXIbus backplane. You may want to
remove one or more of these jumpers to reduce common mode noise.
Figure 1-8. E1468A 8 x 8 Matrix Switch Terminal Module
Column
Connector
Connector
Row
Column Expansion
Connector
*
Column Input
Connector
Row Input
Connector
Row Expansion
Connector
Shield
Connector
TB5
***
In parallel with the screw terminals.
*