Service manual
Test 2-2: DC
Isolation Test
This test verifies that sufficient DC isolation exists at various points on the
Module. DC Isolation is checked from HI to LO, HI to Chassis, and LO to
Chassis. This test uses the test fixture (see Figures 2-1, 2-2, or 2-3).
NOTE
The DMM used should be capable of measuring at least 1 G
Ω.. If the DMM
indicates an overload, record the reading as >Rmax, where Rmax is the
highest resistance that the DMM can measure. For example, if the DMM is
an Agilent 3458A, a typical return for an overload is 1.E+38 and the entry
in Table 2-1 should be >1.2 G
Ω..
HI to LO Isolation 1. Make hardware connections as shown in Figure 2-6
2.
Set DMM to 2-wire ohms, 1 G Ω range
3. Send CLOS (@10000:1rrcc) to the matrix to close all relays
where rr = highest row number and cc = highest column number
For E1465A, rr = 15 and cc = 15
For E1466A, rr = 03 and cc = 63
For E1467A, rr = 07 and cc = 31
4. Trigger the DMM with TRIG SGL
Figure 2-6. HI to LO DC Isolation Test Connections
Agilent E1465A/66A/67A Service Manual Verification Tests 29