Service manual

2. Ensure all relays are open
Send *RST to Module
Trigger the DMM with TRIG SGL and note reading
DMM reading should indicate an open condition (> 1.2 G )
3. Measure Column 00 to Row 00 HI path resistance
Send CLOS (@nn0000) to close column 0 row 0, where nn = card
# (typically 01)
Trigger the DMM with TRIG SGL and note reading
Send OPEN (@nn0000) to open column 0 row 0, where nn = card
# (typically 01)
Enter the result in Table 2-1 for column 0 row 0
4. Repeat for all Columns
Repeat step 3 for all columns
Use CLOS (@nn00cc) and OPEN (@nn00cc), where nn = card #
and cc= column # (omit leading zeroes in nn)
For E1465A use column numbers 01 through 15
For E1466A use column numbers 01 through 63
For E1467A use column numbers 01 through 31
5. Repeat for all Rows
Repeat steps 3 and 4 for all rows
Use CLOS (@nnrrcc) and OPEN (@nnrrcc), where nn = card #,
rr = row #, and cc= column # (omit leading zeroes in nn)
For E1465A use row numbers 01 through 15
For E1466A use row numbers 01 through 03
For E1467A use row numbers 01 through 07
Column LO to Row LO
Measurements 1. Make Hardware Connections
Turn mainframe power OFF
Set DMM to measure 4-wire Ohms
Connect DMM leads as shown in Figure 2-5
Turn mainframe power ON
2. Repeat steps 2 through 5 of the "Column HI to Row HI
Measurements"
24 Verification Tests Agilent E1465A/66A/67A Service Manual