Self-Test
Error Codes
Table 4-4 shows the self-test error codes for the Switch Module. The
meaning of each code is given in the right-hand column. If a self-test failure
occurs, cycle power and repeat the test. If the problem reoccurs, the device
may need repair.
Table 4-4. Self-test Error Codes
Error* Description
+0
+ss01
+ss02
+ss03
+ss10
+ss11
Self-test passes
Firmware error
Bus error (communications problem with card)
Bad ID information in ID register
Interrupt expected but not received
Busy bit was not held 10.5 to 18.5 msec
*ss = card number (with leading zero deleted)
Disassembly Use the following procedures to disassemble the Agilent E1463A Switch
Module. For disassembly refer to Figure 4-1.
Switch Module
Disassembly 1. To remove the covers:
Remove the nine T10 Torx screws from the top cover as shown.
Lift the top cover off of the module.
Turn the assembly over and lift off the bottom cover.
2. To remove the front panel:
Remove the two T8 Torx screws from the front panel handles as
shown.
Remove the T8 Torx screw holding the panel to the A1 PC
Assembly as shown.
Lift the front panel from the A1 assembly.
Chapter 4 Service 49