User`s manual
5. Select the Arbitrary Waveform Function
[SOURce:]FUNCtion[:SHAPe] USER
This command selects the arbitrary waveform function. Couple the
command to the previous frequency command.
6. Set the Maximum Output Amplitude
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <amplitude>
This command specifies the maximum output amplitude. The
amplitude must be equal or greater than the maximum voltage value
of the waveform segment. Refer to Table B-4 in Appendix B for the
amplitude limits.
7. Name the First Waveform Segment
[SOURce:]LIST[1][:SEGMent]:SELect <name>
This command names the first waveform segment.
8. Set the First Waveform Segment Size
[SOURce:]LIST[1][:SEGMent]:DEFine <length>
This command defines the size of the selected waveform segment.
9. Store the First Waveform Segment as Voltages
[SOURce:]LIST[1][:SEGMent]:VOLTage <voltage_list>
This command stores the first waveform segment into the AFG’s
segment memory.
10. Name the Second Waveform Segment
[SOURce:]LIST[1][:SEGMent]:SELect <name>
This command names the second waveform segment.
11. Set the Second Waveform Segment Size
[SOURce:]LIST[1][:SEGMent]:DEFine <length>
This command defines the size of the selected waveform segment.
12. Store the Second Waveform Segment as Voltages
[SOURce:]LIST[1][:SEGMent]:VOLTage <voltage_list>
This command stores the second waveform segment into the AFG’s
segment memory.
13. Name the Segment Sequence
[SOURce:]LIST[1]:SSEQuence:SELect <name>
This command names the segment sequence. The name must be
different from any segment names stored in memory.
14. Set the Segment Sequence Length
[SOURce:]LIST[1]:SSEQuence:DEFine <length>
This command defines the length of the selected segment sequence.
The length must be equal or greater than the number of the waveform
segments stored in memory.
94 Generating Arbitrary Waveforms Chapter 3