Table B-6. Agilent E1445A Error Messages (continued)
Code Message Description
-312 PUD memory lost The protected user data saved by the *PUD
command has been lost.
-313 Calibration memory lost The nonvolatile calibration data used by the *CAL
command has been lost.
-330 Self-test failed Note the information associated with the message for
a description of the failure.
-350 Too many errors The E1445A error queue is full and additional errors
have occurred.
-410 Query INTERRUPTED The E1445A was sent a command before it was
finished responding to a query command.
-420 Query UNTERMINATED The controller (computer) attempts to read a query
response from the Agilent E1445A without having first
sent a complete query command.
-430 Query DEADLOCKED The E1445A’s input and output buffers are full and
the AFG cannot continue.
-440 Query UNTERMINATED after
indefinite response
Occurs when the *IDN? query is not the last query
executed in a command string.
+1000 Out of memory The E1445A segment or sequence memory is full.
+1002 Calibration security enabled Calibration security must be disabled to calibrate the
E1445A, to read or write calibration data, to change
the security code, or to change the protected user
data.
+1004 Calibration write fail Writing calibration or protected user data (*PUD) to
nonvolatile memory failed.
+1005 Calibration constant out of range Illegal calibration constant was computed.
+1006 Calibration constant conflict Calibration constants used during calibration set an
illegal condition.
+1007 Calibration security defeated CALibration secure state disabled and detected at
power-on.
+1011 Illegal while download enabled or
testing local bus
Commands such as SOUR:LIST1 ... cannot be
executed under current conditions. Execute
SOUR:ARB:DOWN:COMP to disable downloading,
or VINS:CONF:TEST:DATA?, to complete the local
bus test.
+1012 Illegal when not downloading SOUR:ARB:DOWN:COMP disables downloading
only after it has previously been enabled.
+1013 Illegal when not testing local bus VINS:CONF:TEST:DATA? was executed and the
local bus test was not performed.
+1014 Illegal while initiated Command cannot be executed while the E1445A is in
the initiated (instrument action) state.