User`s manual

Table B-6. Agilent E1445A Error Messages (continued)
Code Message Description
-161 Invalid block data The number of bytes in a definite length data block
does not equal the number of bytes indicated by the
block header.
-168 Block data not allowed Block data was specified when another parameter
type (i.e. discrete, numeric, boolean) is required.
-178 Expression data not allowed The parameter was specified as an expression
(e.g. SOUR:FREQ1:FIX (A*B).
-183 Invalid inside macro definition Voltage or segment list is inside a macro.
-211 Trigger ignored Trigger was received and the AFG was not in the
wait-for-trigger state. Or, a trigger was received from
a source other than the specified source.
-212 Arm ignored Arm was received and the AFG was not in the
wait-for-arm state. Or, an arm was received from a
source other than the specified source.
-213 Init ignored INITiate:IMMediate received while the AFG was
currently initiated.
-221 Settings conflict See "Settings Conflict Error Messages" at the end of
this table.
-222 Data out of range Parameter value is out of range for any AFG
configuration (e.g. SOUR:FREQ1:FIX 1E9).
-224 Illegal parameter value The calibration security code required to disable
calibration security does not match the stored code.
-241 Hardware missing Command was intended for the Agilent E1446A
which was not present, or is outside the servant area
of the Agilent E1445A AFG.
-270 Macro error
*RMC <
name> was executed and name is not defined.
-272 Macro execution error Macro program data sequence could not be executed
due to a syntax error within the macro definition.
-273 Illegal macro label The macro label defined in the *DMC command was
too long, the same as a common command keyword,
or contained invalid header syntax.
-276 Macro recursion error A macro program data sequence could not be
executed because the sequence leads to the
execution of a macro being defined.
-277 Macro redefinition not allowed A macro label in the *DMC command could not be
executed because the macro label was already
defined.
476 Useful Tables Appendix B