User`s manual
*RCL
*RCL <number> restores a previously stored programming state from one of the
10 possible stored state areas. The
<number> indicates which of the stored state
areas should be used.
This command affects the same command settings as does
*RST. Notable
exceptions include the DAC code format (signed vs. unsigned), the
[SOURce:]LIST
commands, including waveform segment, segment sequence, and frequency list
definitions, the
STATus subsystem commands, and the CALibration:SECure
command state.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
number>
numeric 0 through 9 none
Comments • Executable when Initiated: No
• Coupling Group: None
• Related Commands: *LRN?, *RST, *SAV
•
*RST Condition: all saved states set to the same state as the *RST state
*RMC
*RMC <name> purges only the specified macro definition.
NOTE: At printing time,
*RMC is a command proposed and accepted for a revision
and re-designation of IEEE-488.2.
Comments • Use the *PMC command to purge all macro definitions in one command.
• Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: *DMC, *PMC
•
*RST Condition: None