*PUD and *PUD?
*PUD <data> stores the specified data in the Agilent E1445A’s non-volatile
calibration memory. The data must be sent in IEEE-488.2 definite or indefinite
block format. Calibration security must have been previously disabled.
*PUD? returns the current protected user data in IEEE-488.2 definite block format.
The query form may be executed regardless of the state of calibration security.
Note When shipped from the factory, the protected user data area contains information
regarding when the Agilent E1445A was last calibrated.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mask>
block data
or string
0 through 63 characters none
Comments Executable when Initiated: Yes
Coupling Group: None
*RST Condition: Unaffected
Power-On Condition: Unaffected
Example Setting the Protected User Data
*PUD #17Unit #5 Sets data to “Unit #5”.