User`s manual

Power-On Condition: No macros are defined
Example Define Macro to Restart Waveform
*DMC "RESTART",#19ABOR;INIT Defines macro.
*EMC and *EMC?
*EMC <enable> enables and disables macro usage. When <enable> is zero, macro
usage is disabled. Any non-zero value in the range of -32768 to +32767 enables
macro usage.
*EMC? returns 1 if macro usage is enabled, 0 if disabled.
Comments Macro definitions are not affected by this command.
Executable when Initiated: Yes
Coupling Group: None
*RST Condition: Macro usage is disabled
Power-On Condition: Macro usage is enabled
*ESE and *ESE?
*ESE <mask> enables one or more event bits of the Standard Event Status Register
to be reported in bit 5 (the Standard Event Status Summary Bit) of the Status Byte
Register. The
<mask> is the sum of the decimal weights of the bits to be enabled.
*ESE? returns the current enable mask.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mask>
numeric 0 through 255 none
A 1 in a bit position enables the corresponding event; a 0 disables it.
Comments Executable when Initiated: Yes
Coupling Group: None
Related Commands: *ESR?, *SRE, *STB?
*RST Condition: Unaffected
Chapter 8 Command Reference 417