[:CONFigure]:TEST:DATA?
VINStrument[:CONFigure]:TEST:DATA? returns the received VXIbus Local Bus test
data. The data is returned in 16-bit integer format in an IEEE-488.2 definite block.
Comments Executable when Initiated: No
Coupling Group: None
Related Commands: VINStrument[:CONFigure]:TEST:CONFigure
*RST Condition: None
Power-On Condition: Local Bus testing not configured
Example Testing Local Bus Operation
VINS:CONF:TEST:CONF 100 Configure for 100 byte test
send data
VINS:CONF:TEST:DATA? Read back test data.
[:CONFigure]:VME[:MODE]
VINStrument[:CONFigure]:VME[:MODE] <mode> selects the operating mode for
the VXIbus data transfer bus. The only available mode is
CONSume.
There is no need to send this command since there is only one available mode. The
command is only included for compatibility with the Agilent Virtual
Instrument/Local Bus System Specification.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode>
discrete CONSume none
Comments Executable when Initiated: Yes
Coupling Group: None
*RST Condition: VINStrument:CONFigure:VME:MODE CONSume
Example Setting the VXIbus Data Transfer Bus Operation Mode
VINS:VME CONS Sets CONSume mode.
VINStrument