User`s manual

*RST Condition: VINStrument:CONFigure:LBUS:MODE:AUTO ON
Example Uncoupling Local Bus Operation Mode
VINS:CONF:LBUS:AUTO OFF Uncouple operation mode.
[:CONFigure]:TEST:CONFigure
VINStrument[:CONFigure]:TEST:CONFigure <length> configures the
Agilent E1445A for Local Bus testing. The
<length> parameter indicates that,
during the test, that number of bytes will be sent to the Agilent E1445A. The data
will be placed into unused waveform segment memory. When all data has been
sent, use the
VINStrument[:CONFigure]:TEST:DATA? query to retrieve what the
Agilent E1445A received.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
length> numeric see below none
The valid range for <
length> is 2 through the size of largest available contiguous piece of
waveform segment memory in bytes (2 bytes per point). <
length> must be an even number.
MINimum and MAXimum cannot be used with this command.
Comments Executable when Initiated: Query form only
Coupling Group: None
Related Commands: VINStrument[:CONFigure]:TEST:DATA?
*RST Condition: None
Power-On Condition: Local Bus testing not configured
Example Testing Local Bus Operation
VINS:CONF:TEST:CONF 100 Configures for 100 byte test.
send data
VINS:CONF:TEST:DATA? Reads back test data.
VINStrument
Chapter 8 Command Reference 405