User`s manual

With VINStrument[:CONFigure]:LBUS[:MODE]:AUTO ON set, the Local Bus
operation mode is automatically set to
CONSume when downloading segment or
segment sequence data (
[SOURce:]ARBitrary:DOWNload LBUS command),
directly driving the main output DAC (
[SOURce:]ARBitrary:DAC:SOURce LBUS
command), or providing phase deviation data ([SOURce:]PM:SOURce LBUS
command); the mode is set to OFF when none of these are active.
Executing the
VINStrument[:CONFigure]:LBUS[:MODE] command sets
VINStrument[:CONFigure]:LBUS[:MODE]:AUTO OFF.
Executable when Initiated: Yes
Coupling Group: None
Related Commands: [SOURce:]ARBitrary:DAC:SOURce,
[SOURce:]ARBitrary:DOWNload, [SOURce:]PM:SOURce,
VINStrument[:CONFigure]:LBUS[:MODE]:AUTO
*RST Condition: VINStrument:CONFigure:LBUS:MODE OFF
Example Setting the Local Bus Operation Mode
VINS:CONF:LBUS PIP Sets pipeline (pass through) mode.
[:CONFigure]:LBUS[:MODE]:AUTO
VINStrument[:CONFigure]:LBUS[:MODE]:AUTO <mode> indicates whether the
VXIbus Local Bus operation mode should be automatically set to
CONSume when
downloading segment or segment sequence data (
[SOURce:]ARBitrary:DOWNload
LBUS
command), directly driving the main output DAC
(
[SOURce:]ARBitrary:DAC:SOURce LBUS command), or providing phase
deviation data (
[SOURce:]PM:SOURce LBUS command), and set to OFF when
none of these are active. If
AUTO ON is set, the Local Bus operation mode is
changed as needed; if
OFF is set, the mode must be explicitly set by the
VINStrument[:CONFigure]:LBUS[:MODE] command.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode>
boolean OFF | 0 | ON | 1 none
Comments Executable when Initiated: Yes
Coupling Group: None
Related Commands: [SOURce:]ARBitrary:DAC:SOURce,
[SOURce:]ARBitrary:DOWNload, [SOURce:]PM:SOURce,
VINStrument[:CONFigure]:LBUS[:MODE]
VINStrument
404 Command Reference Chapter 8