User`s manual

:SWEep:TIMer
TRIGger:SWEep:TIMer <period> selects the time between frequency values for
sweep or frequency list generation when
TRIGger:SWEep:SOURce is set to TIMer.
This value set by command is coupled to the
[SOURce:]SWEep:TIME command
value by the following equation:
TIME = TIMer * (points - 1)
where points is the
[SOURce:]SWEep:POINts value for frequency sweeps, or the
length of the frequency list for frequency list generation.
When changing the frequency list length when
[SOURce:]FREQuency[1]:MODE
LIST
is set, or the [SOURce:]SWEep:POINts value when any other MODE is set, the
TIME or TIMer value remains the same, depending on which command,
[SOURce:]SWEep:TIME or TRIGger:SWEep:TIMer respectively, was most recently
sent. The other value is changed based on the new points value.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
period> numeric .00125 through 4.19430375 |
MINimum | MAXimum
seconds
MINimum selects 1.25 mS; MAXimum selects 4.19430375 S.
The above values bound the valid range for <period>
Comments Executable when Initiated: Query form only
Coupling Group: Frequency
Related Commands: [SOURce:]LIST2:FREQuency, [SOURce:]SWEep:POINts,
[SOURce:]SWEep:TIME, TRIGger:SWEep:SOURce
*RST Condition: SOURce:SWEep:TIME 1 is set; TRIGger:SWEep:TIMer is the
dependent value.
Example Setting the Sweep Advance Period
TRIG:SWE:TIM .1 Sets .1 S per frequency value.
TRIGger
402 Command Reference Chapter 8