User`s manual
• *RST Condition: None
Example Advancing a Frequency Sweep
SWE:STAR 1E3;STOP 10E3 Sets sweep frequency limits.
SWE:POIN 10 Sets 1 kHz steps.
ARM:LAY2:SOUR IMM Sets output to start immediately.
ARM:SWE:SOUR IMM Sets sweep to start immediately.
TRIG:SWE:SOUR HOLD Sets sweep to advance sweep manually.
INIT Initiates trigger system.
TRIG:SWE Advances to next frequency.
:SWEep:LINK
TRIGger:SWEep:LINK <link> selects the internal event that advances a frequency
sweep or list when
TRIGger:SWEep:SOURce is set to LINK. The only defined
internal event to advance a sweep or list is “
ARM[:STARt|:SEQuence[1]]:LAYer2”.
There is no need to send this command since there is only one defined internal
event. The command is included for SCPI compatibility purposes only.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
link> string “ARM[:STARt|:SEQuence1]:LAYer2” none
Comments • Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: TRIGger:SWEep:SOURce
•
*RST Condition: TRIGger:SWEep:LINK “ARM[:STARt|:SEQuence[1]]:LAYer2”
Example Linking the Sweep Advance Trigger
TRIG:SWE LINK Links sweep advance trigger to start arm.
TRIG:SWE:LINK "ARM:LAY2"
TRIGger
400 Command Reference Chapter 8