User`s manual

:STOP:SLOPe
TRIGger:STOP:SLOPe <edge> selects the edge (rising or falling) on the
Agilent E1445A’s front panel “Stop Trig/FSK/Gate In” BNC which terminates the
current start arm cycle at the end of the current waveform repetition. This edge is
significant only with
TRIGger:STOP:SOURce set to EXTernal. The programmed
value is retained but not used when other sources are selected.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
edge>
discrete NEGative | POSitive none
Comments Executable when Initiated: Query form only
Coupling Group: Frequency
Related Commands: TRIGger:STOP:SOURce
*RST Condition: TRIGger:STOP:SLOPe POSitive
Example Setting the Stop Trigger Slope
TRIG:STOP:SLOP NEG Sets negative stop trigger slope.
:STOP:SOURce
TRIGger:STOP:SOURce <source> selects the source that can terminate the
current start arm cycle at the end of the current waveform repetition. When the
Agilent E1445A receives a stop trigger, the start trigger sequence is placed into the
wait-for-arm state at the end of the current waveform repetition, aborting the
remaining
ARM[:STARt][:LAYer[1]]:COUNt repetitions of the current arm cycle.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
source> discrete BUS | EXTernal | HOLD | TTLTrg0
through TTLTrg7
none
Comments The available sources are:
BUS: The Group Execute Trigger (GET) GPIB command or the IEEE-488.2
*TRG common command.
EXTernal: The Agilent E1445A’s front panel “Stop Trigger/FSK/Gate In”
BNC connector.
HOLD: Suspend stop triggering. Use the TRIGger:STOP[:IMMediate] command
to terminate a start arm cycle.
TTLTrg0 through TTLTrg7: The VXIbus TTL trigger lines.
TRIGger